Email: info@newmeditech.com | Call Us at +919677296252
X

Product Enquiry

    Tissue Culture Room MTCR

    Tissue Culture Room MTCR

    The Tissue Culture Room MTCR walk-in room provides a multi-tier modular platform available in increments of 60ft2 (5.6m2). This makes the Tissue Culture Room MTCR chamber suitable for research involving plant tissue culture and is often used within large micro propagation programs. The solid side wall plenums transfer air to individual air shelves ensuring upward airflow from each individual shelf, which is essential in mitigating condensation within media dishes.

     

     

    Specification Of Tissue Culture Room MTCR
    Temperature Range +10°C to +45°C lights on
    Growth Area 60 ft2 (6.6m2)
    Growth Height
    20” (510mm) Tissue Culture Room
    External Dimensions 10’7.5” x 5’10” x 9’2.25” (3240mm x 1780mm x 2800mm)
    Internal Capacity 99 ft3 (2820l)
    Electrical Service 120/208-3Ø-60Hz (220/380-3Ø-50Hz)
    Light Intensity 190 μmoles/m2/s
    Weight 4610lb (2091kg)

     

    Product Overview/Applications Tissue Culture Room
    The Tissue Culture Room MTCR walk-in room provides a multi-tier modular platform available in increments of 60ft2. This makes the Tissue Culture Room MTCR chamber suitable for research involving plant tissue culture and is often used within large micro propagation programs. The solid side wall plenums transfer air to individual air shelves ensuring upward airflow from each individual shelf, which is essential to mitigating condensation within media dishes Tissue Culture Room.

    Lighting – Tissue Culture Room
    The standard lighting package for the Tissue Culture Room MTCR model generates low to moderate light intensities and incorporates fluorescent lamps. Intensities are ideally suited for tissue culture research. Multiple lighting levels can be programmed with the control system while higher light intensities are available as an option.

    Airflow – Tissue Culture Room
    A center column conditioning coil directs air through pressurized sidewall plenums which is then directed into each individual air shelf. Uniform vertical airflow from each shelf ensures that a precise volume of air is directed upward around the media containers. Tissue Culture Room Filtered and adjustable fresh air intake and exhaust openings enable researchers to manage exchange air to the chamber.

    Refrigeration – Tissue Culture Room
    The standard refrigeration system for the Tissue Culture Room MTCR incorporates a water-cooled condensing unit for heat exchange and an electronic modulating valve provides tight temperature control while ensuring quiet operation.Tissue Culture Room Pressure transducers are included for monitoring the status of the refrigeration system Alternative refrigeration methods are available for this equipment platform depending on site specific and/or user defined requirements. Consult the factory for heat rejection information and other options for cooling Tissue Culture Room.

    Experiment Protection – Tissue Culture Room
    User programmable “set and forget” alarms track the chamber’s operation versus user-defined set points. This allows for exceptionally accurate monitoring without the need for adjustment every time the set point is redefined. Backup “high/low” alarms provide a further level of protection while visual and audible notification is provided when any alarm is activated. Contacts for connection to a building management system are also included.

    Key Product Attributes
    • Multi-tier design maximizes tissue culture area
    • Pressurized sidewall plenums deliver uniform vertical airflow to mitigate or eliminate condensation
    • Tissue Culture Room Product certifications/markings : CE